FEATURES

  • 4-20 mA Input
  • Modern, Sleek and Practical Enclosure
  • 3½ Digits, 1" (25.4 mm) Display
  • Display Mountable at 0°, 90°, 180°, & 270° Degrees
  • Easy Calibration and Installation
  • Loop-Powered Backlight Option
  • 1 V Drop (4 V with Backlight)
  • Explosion-Proof, IP68, NEMA 4X Enclosure
  • Flanges for Wall or Pipe Mounting
  • HART® Protocol Transparent
  • Operates from -40 to 75°C
Protex Launch Guide and Product Comparison
Hazardous Area Approvals for PD663 ProTex.
 

INTRODUCTION

The PD6870 is a rugged, loop-powered meter with 1" display digits in an explosion-proof enclosure for demanding applications in hazardous areas or in the harshest environmental conditions. The meter derives all of its power from the 4-20 mA loop; with a small
1 volt drop for use in almost any current loop and application.

It is programmed using four easy to access front-mounted trim potentiometers. No complex or difficult to read programming menus are necessary for setup. The numeric display will read from -1999 to 1999 over a 2000 count user adjustable scaling span. The backlight option lets you see the display under any lighting condition and is powered from the 4-20 mA loop with no additional power supply required.

The enclosure has two threaded conduit holes, one pre-installed conduit plug, and integrated pipe or wall mounting slotted flanges.

 

CONNECTIONS

PD6800 Input Connections with Loop-Powered Backlight

PD6870 Input Connections with Loop-Powered Backlight

 

DIMENSIONS

PD6870 Mounting Dimensions
 

SPECIFICATIONS

Except where noted all specifications apply to operation at +25°C.

General

Display: 3½ digit LCD, 1" (25.4 mm); -1,999 to 1,999
Display Update Rate: 2.5 Updates/Second
Display Orientation: Display may be mounted at 90° increments up to 270° from default orientation
Overrange: Display reads 1 on the left most digit
Programming Method: Four front mounted trim potentiometers accessed when the cover is removed.
Recalibration: Recommended at least every 12 months.
Normal Mode Rejection: 60 dB rejection ratio
Operating Temperature Range: -40 to 75°C
Storage Temperature Range: -40 to 75°C
Relative Humidity: 0 to 90% non-condensing
Connections: Screw terminals accept 12 to 22 AWG wire
Enclosure: Explosion-proof die cast aluminum with glass window, 0.30% max copper content, corrosion resistant epoxy coating, color: blue. NEMA 4X, IP68. Two ¾" NPT threaded conduit openings. One ¾" NPT nickel plated brass conduit plug with 10 mm hex key fitting installed.
Mounting: May be mounted directly to conduit. Two slotted flanges for wall mounting or NPS 1½" to 2½" or DN 40 to 65 mm pipe mounting.
Overall Dimensions: 5.65" x 5.25" x 4.86" (144 mm x 133 mm x 124 mm) (W x H x D)
Weight: 5.00 lbs (80 oz, 2.27 kg)
Warranty: 3 years parts and labor

Input

Input: 4-20 mA
Accuracy: ±0.1% of full span ±1 count
Temperature Drift: 150 PPM/°C from -40 to 75°C ambient
Decimal Point: User selectable decimal point
Calibration Range: 4 mA input: -1000 to +1000; 20 mA between 20 and 2000 counts greater than 4 mA display. Two point linear display span.
Maximum Voltage Drop: 1 VDC @ 20 mA with backlight; 4 VDC @ 20mA with loop-powered backlight
Input Overload: Over current protection to 2 A max.
Loop-Powered Backlight Option: Factory installed only. Powered directly from the 4-20 mA loop, no batteries required. Backlight can be enabled or disabled via alternative wiring of terminal block. The display brightness will increase as the input signal current increases..

Product Ratings & Approvals

FM: Explosion-proof for use in Class I, Division 1, Groups B, C, D. Class II, Division 1, Groups E, F, G. Class III, Division 1; T6. Class I, Zone 1, AExd IIC T6 Gb. Zone 21, AEx tb IIIC T85°C. Ta = -40 to 75°C.
Enclosure: Type 4X & IP66. Certificate number: 3040391
ATEX: II 2 G D. Ex d IIC T6 Gb. Ex tb IIIC T85°C Db IP68. Ta = -40 to 75°C.
Certificate number: Sira 10ATEX1116X
IECEx: IECEx SIR 10.0056X. Ex d IIC T6 Gb. Ex tb IIIC T85 C Db IP68. Ta = -40 to 75°C
CSA: Class I, Division 1, Groups B, C, D. Class II, Division 1, Groups E, F, G. Class III, Division 1; T6. Class I, Zone 1, Ex d IIC T6.
Ta = -40 to 75°C. Enclosure: Type 4X & IP66.
Certificate number: 11 2325749
 

ORDERING INFORMATION

Products

Model Description Price
PD6870-0L0        Loop-Powered Ex-Proof Process Meter: No Options Installed        479.00 Buy Now
PD6870-0K0        Loop-Powered Ex-Proof Process Meter with Backlight        529.00 Buy Now
  

Accessories

Model Description Price
PDAPLUG75        Metal Stopping Plug with Approvals        20.00
PDA0001        ¾" M-NPT to M20 Approved Reducer for the PD6800, PD6820, PD6830 and PD6870        25.00
PDA0002        ¾" M-NPT to ½" F-NPT Approved Reducer for the PD6800, PD6820, PD6830 and PD6870        25.00
PDA6846        Pipe Mounting Kit for PD6800, PD6820, PD6830, PD6870 and PD8 (PD8 requires 2). Includes zinc-plated u-bolt for 2" pipe, (2) washers, and (2) nuts.        5.00
PDA6846-SS        Pipe Mounting Kit for PD6800, PD6820, PD6830, PD6870 and PD8 (PD8 requires 2). Includes zinc-plated u-bolt for 2" pipe, (2) washers, and (2) nuts.        16.00


 

DOCUMENTATION

Precision Digital Return User Form

Previously registered users of the site can access documentation by entering their email address and password below and selecting Login. Unregistered users can access documentation by selecting New User and entering registration information.

Email:
Password:



Add New User Information Forgot Password




Precision Digital Corporation Since 1974 ISO 9001 Certified
Precision Digital Site Map Precision Digital Privacy Policy Precision Digital Copyright 2000-2011